Education, tips and tricks to help you conduct better fMRI experiments.
Sure, you can try to fix it during data processing, but you're usually better off fixing the acquisition!

Tuesday, December 16, 2014

Updated checklist for fMRI acquisition methods reporting in the literature


This post updates the checklist that was presented back in January, 2013. The updated checklist is denoted Version 1.2. The main update is to include reporting for simultaneous multi-slice (SMS) (a.k.a. multi-band, MB) EPI.

Explanatory notes for parameter names appear in the lower portion of the post. Note that the present checklist was devised by considering typical fMRI experiments conducted on 1.5 T and 3 T scanners but the list should work reasonably well for 7 T sequences, too.

Please keep the comments and feedback coming. This is an ongoing, iterative process.




Release notes for Version 1.2

All changes from Version 1.1 have been highlighted in yellow, both on the list PDF and on the explanatory notes (below).

1. The "Spatial Encoding" parameter categories have been renamed "In-Plane Spatial Encoding" to better differentiate in-plane acceleration (e.g. GRAPPA) from slice dimension acceleration (SMS/MB).

2. When using slice dimension acceleration (i.e. SMS/MB), certain parameters that are listed as Supplemental for other EPI variants should be considered Essential. Specifically, it is suggested to report:
Matrix coil mode
Coil combination method

All the In-Plane Spatial Encoding parameters in the Supplemental category should be considered because there is a tendency to use SMS/MB to attain high spatial resolution, requiring long readout echo trains that can have higher distortion than found in typical EPI scans.

The In-plane reconstructed matrix parameter should be reported whenever partial Fourier sampling is used, as it often is for SMS/MB EPI.

All the RF & Contrast parameters in the Supplemental category should be reported because the shape, duration and amplitude of the excitation RF pulse are all integral components of the acceleration method.

The Shim routine should be reported if a non-standard shim is performed before SMS/MB EPI.

3. Pre-scan normalization has been added to the Supplemental section of RF & Contrast parameters. Large array coils produce strong receive field heterogeneity and the use of pre-scan normalization may improve the performance of post hoc motion correction.