Starting points for SMS-EPI at 3 T: Part II
In an earlier post I presented three starting protocols for the CMRR version of SMS-EPI, referred to as the MB-EPI sequence here. I'll use italics to indicate a specific pulse sequence whereas SMS-EPI, no italics, refers to the family of simultaneous multi-slice methods. In this post I'll develop a similar set of three starting protocols for the Massachusetts General Hospital (MGH) version of SMS-EPI, called Blipped-CAIPI. I'm going to build upon the explanations of the last post so please cross reference for parameter explanations and background.
As for the previous post there are several things to bear in mind. This series is Siemens-centric, specifically Trio-centric. While many of the concepts and parameter options may apply to other platforms there will be minor differences in parameter naming conventions and, perhaps, major differences in implementation that you will need to consider before you proceed. For Siemens users, I am running aging software, syngoMR version B17A. The age of the software and the old reconstruction board on the scanner means that you can expect to see much faster reconstruction on a newer system. I hope, but cannot guarantee, that the actual image quality and artifact level won't differ massively from a Trio running VB17A to a new Prisma running VE11C. I'll keep you updated as I learn more.
Preliminaries
As before, for this post I am going to be using a 32-channel receive-only head coil. The SMS-EPI sequences can be made to work with a 12-channel coil but only in a reduced fashion because the 12-channel coil has minimal receive field heterogeneity along the magnet z axis - the struts run parallel with the magnet axis except at the coil's rear, where they converge - and generally we want to do axial slices (along z) for fMRI. I don't yet know whether SMS-EPI would work well on the 20-channel head/neck coil on a Prisma, it's something I hope to investigate in the near future. But a 64-channel head/neck coil on a Prisma will definitely work for SMS-EPI. Better or worse than a 32-channel coil on a Prisma? I have no idea yet.
The Blipped-CAIPI sequence version 2.2 was obtained through a C2P (Core Competence Partnership) with MGH. Installation was a breeze: a single executable to port to the scanner and one click, done. The development team offers an informative but brief 7-page manual which will be useful to anyone who has read the SMS-EPI literature and has a basic understanding of how SMS works. It's not a starting point for everyday neuroscience, however. The manual mentions a .edx (protocol) file as a starting point for 2, 2.5 and 3 mm resolution scans, but in the file I downloaded for VB17A the contents didn't include it. Perhaps contact MGH if you are on another software version and you'd like a .edx file rather than building your own protocol, e.g. by recreating what you see here.
Tuesday, February 9, 2016
Starting points for SMS-EPI at 3 T
Options for SMS-EPI on a Siemens 3 T scanner
I am aware of three SMS-EPI pulse sequences for a Siemens Trio. One comes from the University of Minnesota's Center for Magnetic Resonance Research (hereafter CMRR), one comes from the Martinos Center for Biomedical Imaging at Massachusetts General Hospital (hereafter MGH), and one comes from Siemens as a work-in-progress (WIP) aftermarket sequence. For this post I'm going to be using the sequence provided by CMRR. Since CMRR refer to their sequence as multiband (MB) EPI I shall stick to this nomenclature here, and reserve the term SMS-EPI to apply to the broader family of pulse sequences. I may do posts on the MGH and WIP sequences in the future, but the CMRR sequence has been used the most broadly to date (e.g. the Human Connectome Project, which I'll discuss at length below) and so it offers the most immediate, road-tested place to start.
Tuesday, January 28, 2014
Partial Fourier versus GRAPPA for increasing EPI slice coverage
This is the final post in a short series concerning partial Fourier EPI for fMRI. The previous post showed how partial Fourier phase encoding can accelerate the slice acquisition rate for EPI. It is possible, in principle, to omit as much as half the phase encode data, but for practical reasons the omission is generally limited to around 25% before image artifacts - mainly enhanced regional dropout - make the speed gain too costly for fMRI use. Omitting 25% of the phase encode sampling allows a slice rate acceleration of up to about 20%, depending on whether the early or the late echoes are omitted and whether other timing parameters, most notably the TE, are changed in concert.
But what other options do you have for gaining approximately 20% more slices in a fixed TR? A common tactic for reducing the amount of phase-encoded data is to use an in-plane parallel imaging method such as SENSE or GRAPPA. Now, I've written previously about the motion sensitivity of parallel imaging methods for EPI, in particular the motion sensitivity of GRAPPA-EPI, which is the preferred parallel imaging method on a Siemens scanner. (See posts here, here and here.) In short, the requirement to obtain a basis set of spatial information - that is, a map of the receive coil sensitivities for SENSE and a set of so-called auto-calibration scan (ACS) data for GRAPPA - means that any motion that occurs between the basis set and the current volume of (accelerated) EPI data is likely to cause some degree of mismatch that will result in artifacts. Precisely how and where the artifacts will appear, their intensity, etc. will depend on the type of motion that occurs, whether the subject's head returns to the initial location, and so on. Still, it behooves us to check whether parallel imaging might be a better option for accelerating slice coverage than partial Fourier.
Deciding what to compare
Disclaimer: As always with these throwaway comparisons, use what you see here as a starting point for thinking about your options and perhaps determining your own set of pilot experiments. It is not the final word on either partial Fourier or GRAPPA! It is just one worked example.
Okay, so what should we look at? In selecting 6/8ths partial Fourier it appears that we can get about 15-20% more slices for a fixed TR. It turns out that this gain is comparable to using GRAPPA with R=2 acceleration with the same TE. To keep things manageable - a five-way comparison is a sod to illustrate - I am going to drop the low-resolution 64x48 full Fourier EPI that featured in the last post in favor of the R=2 GRAPPA-EPI that we're now interested in. For the sake of this comparison I'm assuming that we have decided to go with either pF-EPI or GRAPPA, but you should note that the 64x48 full Fourier EPI remains an option for you in practice. (Download all the data here to perform for your own comparisons!)
I will retain the original 64x64 full Fourier EPI as our "gold standard" for image quality as well as the two pF-EPI variants, yielding a new four-way comparison: 64x64 full Fourier EPI, 6/8pF(early), 6/8pF(late), and GRAPPA with R=2. Partial Fourier nomenclature is as used previously. All parameters except the specific phase encode sampling schemes were held constant. Data was collected on a Siemens TIM/Trio with 12-channel head coil, TR = 2000 ms, TE = 22 ms, FOV = 224 mm x 224 mm, slice thickness = 3 mm, inter-slice gap = 0.3 mm, echo spacing = 0.5 ms, bandwidth = 2232 Hz/pixel, flip angle = 70 deg. Each EPI was reconstructed as a 64x64 matrix however much actual k-space was acquired. Partial Fourier schemes used zero filling prior to 2D FT. GRAPPA reconstruction was performed on the scanner with the default vendor reconstruction program. (Siemens users, see Note 1.)
Thursday, December 19, 2013
Using partial Fourier EPI for fMRI
Back in August I did a post on the experimental consequences of using partial Fourier for EPI. (An earlier post, PFUFA Part Fourteen introduces partial Fourier EPI.) The main point of that post was to demonstrate how, with all other parameters fixed, there are two principal effects on an EPI obtained with partial Fourier (pF) compared to using full phase encoding: global image smoothing, and regionally enhanced signal dropout. (See Note 1.)
In this post I want to look a little more closely at how pF-EPI works in practice, on a brain, with fMRI as the intended application, and to consider what other parameter options we have once we select pF over full k-space. I'll do two sets of comparisons. In the first comparison all parameters except the phase encoding k-space fraction will be fixed so that we can again consider the first stage consequences of using pF. In the second comparison each pF-EPI scheme will be optimized in a "maximum performance" test. The former is an apples to apples comparison, with essentially one variable changing at a time, whereas the latter is how you would ordinarily want to consider the pF options available to you.
Why might we want to consider partial Fourier EPI for fMRI anyway?
If we assume a typical in-plane matrix of 64 x 64 pixels, an echo spacing (the time for each phase-encoded gradient echo in the train, as explained in PFUFA Part Twelve) of 0.5 ms and a TE of 30 ms for BOLD contrast then it takes approximately 61 ms to acquire each EPI slice. (See Note 2 for the details.) The immediate consequence should be obvious: at 61 ms per slice we will be limited to 32 slices in a TR of 2000 ms. If the slice thickness is 3 mm then the total brain coverage in the slice dimension will be ~106 mm, assuming a 10% nominal inter-slice gap (i.e. 32 x 3.3 mm slices). With axial slices we aren't going to be able to cover the entire adult brain. We will have to omit either the top of parietal lobes or the bottom of the temporal lobes, midbrain, OFC and cerebellum. Judicious tilting might be able to capture all of the regions of primary interest to you, but we either need to reduce the time taken per slice or increase the TR to cover the entire brain.
Partial Fourier is one way to reduce the time spent acquiring each EPI slice. There are two basic ways to approach it: eliminate either the early echoes or the late echoes in the echo train, as described at the end of PFUFA: Part Fourteen. Eliminating the early echoes doesn't, by itself, save any time at all. Only if the TE is reduced in concert is there any time saving. But omitting the late echoes will mean that we complete the data acquisition for the current slice earlier than we would for full Fourier sampling, hence there is some intrinsic speed benefit. I'll come back to the time savings and their consequences later on. Let's first look at what happens when we enable partial Fourier without changing anything else.
Monday, August 5, 2013
The experimental consequences of using partial Fourier for EPI
PFUFA Part Fourteen introduced the idea of acquiring partial k-space and explained how the method, hereafter referred to as partial Fourier (pF), is typically used for EPI acquisitions. At this point it is useful to look at some example data and to begin to assess the options for using pF-EPI for experiments.
Image smoothing
The first consequence of using pF is image smoothing. It arises because we've acquired all of the low spatial frequency information twice - on both halves of k-space - but only half of some of the high spatial frequency information. We've then zero-filled that part of k-space that was omitted. This has the immediate effect of degrading the signal-to-noise ratio (SNR) for the high spatial frequencies that reside in the omitted portion of k-space. (PFUFA Part Eleven dealt with where different spatial frequencies are to be found in k-space.) Thus, the final image has less detail and is smoother than it would have been had we acquired the full k-space matrix, and because of the smoothing the final image SNR tends to be higher for pF-EPI than for the full k-space variant.
It was surprising to me that pF-EPI has higher SNR - due to smoothing - than full Fourier EPI in spite of the reduced data sampling in the acquisition. Conventional wisdom, which is technically correct, states that acquiring less data will degrade SNR. To understand this conundrum, we can think of pF as being like a square filter applied asymmetrically to the phase encoding dimension of an EPI obtained from a complete k-space acquisition. Indeed, as we start to evaluate the costs and benefits of pF for EPI we should probably be thinking about a minimum of a three-way comparison. Firstly, we obviously want to compare our pF-EPI to the full k-space alternative having the same nominal resolution. But we should also consider whether there is any advantage over a lower resolution EPI with full k-space coverage, too. Why? Because this lower resolution version is, in effect, what you get when partial Fourier is applied symmetrically, i.e. when the high spatial frequencies are omitted from both halves of the phase encoding dimension!
Let's do our first assessment of pF on a phantom. There are four images of interest: the full k-space image, two versions of pF - omitting the early or the late echoes from the echo train - and, for the sake of quantifying the amount of smoothing, a lower resolution full k-space image which is tantamount to omitting both the early and late echoes. (See Note 1.) From this point on I'm going to refer to omission of the early and late echo variants as pF(early)-EPI and pF(late)-EPI, respectively.
Wednesday, June 12, 2013
Physics for understanding fMRI artifacts: Part Fourteen
Partial Fourier EPI
(The full contents for the PFUFA series of posts is here.)
In PFUFA Part Twelve you saw how 2D k-space for EPI is achieved in a single shot, i.e. using a repetitive gradient echo series following a single excitation RF pulse. The back and forth gradient echo trajectory permits the acquisition of a 2D plane of k-space in tens of milliseconds. That's fast to be sure, but when one wants to achieve a lot of three-dimensional brain coverage then every millisecond counts.
In the EPI method as presented in PFUFA Part Twelve it was (apparently) necessary to cover - that is, to sample - the entire k-space plane in order to then perform a 2D Fourier transform (FT) and recover the desired image. Indeed, this "complete" sampling requirement was developed earlier, in PFUFA Part Nine, when we looked at 2D k-space and its relationship to image space.
One aspect of the FT that I glossed over in previous posts has to do with symmetry. Perhaps the eagle-eyed among you spotted the symmetry in the 2D k-space of the first couple of pictures in PFUFA Part Nine. If you didn't, don't worry about it because I'm about to show it to you in detail. It turns out that there's actually no need to acquire the entire 2D k-space plane; it suffices to acquire some of it - at least half - and then use post-processing methods to fill in the missing part. At that point one can apply the 2D FT and recover the desired image.
Now, as you would expect, there's no free lunch on offer. There are practical consequences from not acquiring the full k-space plane. In this post we will look briefly at the physical principles of partial Fourier EPI, then in the next post we'll take a look at some example data that will provide a basis for evaluating partial versus full k-space coverage for fMRI.
(The full contents for the PFUFA series of posts is here.)
In PFUFA Part Twelve you saw how 2D k-space for EPI is achieved in a single shot, i.e. using a repetitive gradient echo series following a single excitation RF pulse. The back and forth gradient echo trajectory permits the acquisition of a 2D plane of k-space in tens of milliseconds. That's fast to be sure, but when one wants to achieve a lot of three-dimensional brain coverage then every millisecond counts.
In the EPI method as presented in PFUFA Part Twelve it was (apparently) necessary to cover - that is, to sample - the entire k-space plane in order to then perform a 2D Fourier transform (FT) and recover the desired image. Indeed, this "complete" sampling requirement was developed earlier, in PFUFA Part Nine, when we looked at 2D k-space and its relationship to image space.
One aspect of the FT that I glossed over in previous posts has to do with symmetry. Perhaps the eagle-eyed among you spotted the symmetry in the 2D k-space of the first couple of pictures in PFUFA Part Nine. If you didn't, don't worry about it because I'm about to show it to you in detail. It turns out that there's actually no need to acquire the entire 2D k-space plane; it suffices to acquire some of it - at least half - and then use post-processing methods to fill in the missing part. At that point one can apply the 2D FT and recover the desired image.
Now, as you would expect, there's no free lunch on offer. There are practical consequences from not acquiring the full k-space plane. In this post we will look briefly at the physical principles of partial Fourier EPI, then in the next post we'll take a look at some example data that will provide a basis for evaluating partial versus full k-space coverage for fMRI.