Sure, you can try to fix it during data processing, but you're usually better off fixing the acquisition!

Wednesday, March 21, 2012

GRAPPA and multi-band imaging. And motion. Again.

 

It's come to my attention that some of the latest accelerated (aka multiplexed) EPI sequences are now being made available to some sites with vendor/collaborative research agreements, a move that should catalyze their verification, testing and eventual application for neuroscience. The distribution of these pulse sequences to the wider world is great news! The potential is considerable! However, those wanting to conduct neuroscience experiments today with these zippy new tools should bear in mind the not inconsiderable risks. I want to warn you to think very carefully before taking the plunge.

Today's accelerated EPI sequences combine techniques such as multi-band (MB) acquisition with simultaneous echo refocusing (SER) and/or GRAPPA (1,2). In previous posts I've highlighted the increased motion sensitivity of parallel imaging methods such as GRAPPA. The MB family of methods also require "reference scan data" in order to reconstruct the time series images, and as such they are inherently more motion-sensitive than your plain vanilla single-shot EPI. Indeed, similar principles are used to reconstruct MB images as for GRAPPA, and the basic motion sensitivities are the same, i.e. motion during the reference data acquisitions will contaminate all images in a subsequent time series, while motion after the reference data but during the (accelerated) time series will lead to mismatches and spatial artifacts that will degrade temporal stability. In short, using these accelerated sequences is akin to sharpening the motion sensitivity profile of your experiment, and you will need to ensure a high degree of subject compliance to get good data.


Plan, then scan.

Now, I'm not suggesting you dismiss out of hand these sequences for your research. I am suggesting that you apply a lot of forethought, taking the time to consider several important factors. I've written before about evaluating pulse sequences that are new (or new to you). Your first task is to determine whether you even need a fancy, partly validated, highly risky pulse sequence to answer your neuroscience question. If the answer isn't a resounding "yes," why take the risk? Next, you should ask yourself how the pulse sequence should be set up to provide the optimum data. For instance, do you know which slice direction is best for minimizing motion sensitivity and/or receive field bias (g-factor) for the multi-band sequence? And do you know which RF coil to use, and why? If you can't establish your experimental setup based on sound principles that's a suggestion you either don't have the expertise yourself or you aren't collaborating with someone with the requisite expertise. (Me? I could guess, but that's about it! Without doing a validation study of my own I'd be winging it. Which is kinda my point!)

Please don't just go download and use the latest and greatest technique because it's new and cool. I've seen this movie before, and ninety nine times out of a hundred it ends in tears. Please put some justification and logic into your choices before you go and spend hundreds of hours and thousands of dollars finding yet another way that motion can confound an fMRI experiment. Eyes wide open!

__________________


References:

1.  S Moeller, et al. "Multiband multislice GE-EPI at 7 tesla, with 16-fold acceleration using partial parallel imaging with application to high spatial and temporal whole-brain fMRI." Magn. Reson. Med. 63, 1144-53 (2009).

2.  DA Feinberg, et al. "Multiplexed echo planar imaging for sub-second whole brain fMRI and fast diffusion imaging." PLoS ONE 5(12), e15710 (2010).

Tuesday, March 13, 2012

GRAPPA: another warning about motion sensitivity

 
I wrote a post in May last year to highlight the enhanced motion sensitivity of GRAPPA-EPI compared to single-shot EPI for fMRI. Paul Mullins and I had also discussed the use of GRAPPA for resting-state fMRI in the Comments of an earlier post. The literature is still fairly quiet on the adverse effects of GRAPPA for fMRI although, as I noted in the May post, there are one or two reports of reduced fMRI sensitivity when using parallel imaging, some of which might be attributable to motion (whether it was diagnosed as motion or not in the published work).

In the May, 2011 post I explained the two types of motion sensitivity that plague GRAPPA in its usual incarnation for EPI time series acquisitions. The first type - motion contamination of the auto-calibration scans (ACS) - might be mitigated by vigilance and a suitably resilient task script, e.g. one that uses plenty of null events at the start of the acquisition, before the first real stimulus is presented, to give the operator sufficient time to evaluate the images being generated with the current ACS and decide whether or not to stop and start over. This approach is no guarantee that motion won't have contaminated the ACS, but simple tactics like this can help avoid the worst effects of motion during the start of the run.

The second type of motion is that which happens after the ACS and during the (under-sampled) time series itself. This problem is one of mismatch. Displacement of the head from its position during the ACS acquisition can lead to spatial errors in the current image volume. Thus, whilst attaining motion-free ACS might be considered essential for fMRI, maintaining proper matching of the ACS to the under-sampled time series is also important. The bigger the mismatch the more likely there will be a penalty in statistical power for the time series.

In this post I want to tackle the issue of non-head motion in the scanner, and its effects on GRAPPA-EPI images. This investigation was motivated by one of my users who reported seeing occasional "banding" in a study that had used GRAPPA-EPI. The traditional evaluation of head motion suggested that the subjects weren't moving very much, so I started looking into other possible instabilities. I was quite surprised just how sensitive GRAPPA-EPI can be to small perturbations, as you will shortly see.


A quick review of some brain data

Let's begin by looking at one of the problem GRAPPA-EPI data sets from a human subject. The acquisition specifics are as follows: 12-channel head RF coil on a Siemens Trio/TIM scanner, GRAPPA with R = 2, reconstructed matrix = 96x96, FOV = 224x224 mm, slice thickness = 3 mm, 10% gap, interleaved sagittal slices, flip angle = 90 deg, TR/TE=2000/26 ms, echo spacing = 0.8 ms, readout bandwidth = 1408 Hz/pixel.

Here is a cine-loop through the raw data:


Thursday, March 8, 2012

New user training guide/FAQ

I've just uploaded a new user training guide/FAQ that we use at Berkeley to initiate newbies into the ways of the dark side. It is Siemens-specific, for a Trio/TIM.

As last time, the guide is a bit rough. Sorry for English-isms and typos. It's worth exactly what you pay for it. It's free. Use and abuse it however you like. It's a Word document so that you can reorder things, add your own notes, etc. I would appreciate constructive feedback, especially if you find mistakes or have suggestions to improve it, but there's no need to ask permission to use it, change it, replicate it, sell it...

The most recent version of the training guide/FAQ is available from this web page:

http://bic.berkeley.edu/scanning

Locate the file attachment towards the bottom of the page, it's called 3T_user_training_FAQ_08Mar2012.doc. The most recent contents and a list of changes since the last version (April, 2011) appear below.


Caveat emptor.

The document is only a component of user training, don't expect to learn how to scan by reading it! Rather, use the tips to extend your understanding, refine your experimental technique and so on. Note also that this document is for a Siemens TIM/Trio (with 32 receive channels) and running software VB17. There may be subtle or not-so-subtle differences for the Verio and Skyra platforms, for software VB15, VD11, etc. so keep your wits about you if you're not on a Trio with VB17!

You may have local differences, e.g. custom pulse sequences, that allow you to do things that contradict what you find in this user guide. Talk to your physicist and your local user group before taking anything you find in this guide/FAQ too literally.

Finally, you wont find many (any?) references in this guide/FAQ. It's for the training of newbies, not a comprehensive literature review! If you are seeking further information on something I mention in the guide and you can't find a suitable reference yourself, shoot me an email and I'll do my best to point you in a useful direction.

 --------------------------------------


Update Notes (8th March, 2012):

  • Updated with new operating modes available under software syngo MR version B17.
  • General tweaks to improve readability.
  • Further recommendations on using the 32-channel coil for fMRI.
  • Added a description of the new AutoAlign procedure, AAHScout.
  • Added a new section: “I have an existing protocol that uses the old AutoAlign (AAScout). How do I get and use the new AutoAlign (AAHScout)?”
  • Added a new section: “I want to add a new acquisition and acquire exactly the same slices as this other EPI acquisition I just acquired. How do I tell the scanner to do that?”
  • Extended the discussion on the relative merits of PACE versus using an offline realignment alone, in the section on the ep2d_pace sequence.
  • Fixed a typo concerning the slice ordering for descending slices.
  • Added a new section: “What is a field map and how does it fix EPI distortion?”
  • Added a new section: “I want to try to fix my distortion with a field map. What do I need to acquire?”
  • Updated the sections on partial Fourier for EPI, noting that Siemens simply zero fills the omitted portion of k-space rather than doing a conjugate synthesis.
  • Extended checklists.

Tuesday, February 28, 2012

Common persistent EPI artifacts: Distortion and dropout

The origins of distortion and dropout in EPI were covered in PFUFA Part Twelve, and both of these artifacts have been mentioned in passing in the previous articles concerning abnormally high ghosting. In some instances these artifacts are "co-morbid" because certain issues that cause abnormally high ghosting - such as a poor shim because of asymmetric placement of the subject's head in the magnet - are likely to increase distortion and dropout effects at the same time. Except that it can be very difficult to evaluate distortion and dropout by inspection, during an experiment. The ghosts can be used as a fairly independent "barometer" of the experiment's quality if, as is often the case, some of them fall into an image region that is otherwise noise. Not so with distortion and dropout. By definition these artifacts plague signal regions in the brain, and even an experienced operator can have a tough time determining when either issue is worse than it might otherwise be.

So I'm afraid I don't have a whole lot of new information to offer on either distortion or dropout, from the perspective of diagnosing and potentially changing (improving) your experiment on the day. Other than very obvious deficiencies, as might happen if the subject has a highly conductive hair product, for example, I don't spend much time evaluating distortion or dropout by inspection. Ghosts can be a good surrogate for all that ails distortion and dropout, so I focus on those.

Where you can potentially improve the situation for distortion and dropout is with parameter selection when you are establishing your experimental protocol. Distortion and dropout will generally change with slice prescription, as we already saw in the "good data" posts. And it may be that reduction of dropout leads you to use a particular slice direction, e.g. coronal slices for improved frontal lobe signal. After that, the other common tactics to minimize dropout are to use the thinnest possible slice thickness, possibly using higher in-plane spatial resolution, and perhaps decrease TE. These are protocol/parameter questions that are covered somewhat in my user training guide/FAQ, and I will expand on those sections below. Be warned, however, that it is very difficult to provide general guidelines for all fMRI experiments. Instead, the parameter choices tend to be dictated by your primary requirements. You might select very different parameters for a study that is primarily interested in orbitofrontal cortex than you would use for a sensorimotor task. It's horses for courses.


Approaches to tackling distortion

The level of distortion in the phase encoding dimension is a function of the echo spacing, as explained in PFUFA Part Twelve. Tactics to reduce the distortion level involve making fundamental changes to the phase encoding k-space scheme, e.g. multi-shot segmented k-space, or parallel imaging methods. In each approach the essential idea is to increase the k-space step size, thereby increasing the bandwidth of the phase encoding dimension.

Sunday, February 19, 2012

Terminology change: characterizing EPI artifacts

After considering the rest of the topics that I want to cover in the series of posts on EPI artifacts, I've decided to change the terms "static" and "dynamic" to "persistent" and "intermittent," respectively. I think the new terms better reflect the dominant temporal character of each artifact. The idea is to sort them based on whether they are likely to plague every frame of an EPI time series, or come and go.

Take external RF interference, for example. Say you fail to close properly the RF-sealed door to the magnet room and the door opens slightly during your scan, leading to RF contamination from "environmental" sources. In this instance the RF interference itself isn't likely to be static - it will vary with whatever sources of RF happen to be in your scanner environment - but it will persist (at some level) until the door is closed. With a good eye or some appropriate diagnostics it would be possible to show the persistence of the problem throughout an acquisition. Contrast this situation with a static electrical discharge somewhere within the scanner room; tiny sparks that cause a broad range of electromagnetic frequencies, including radiofrequencies. These can arise if the humidity of the magnet room air becomes too low. Depending on the source of the sparks, the humidity, etc. you might find that only one or two TRs of a time series are contaminated, or you could find the entire time series is affected. I will therefore characterize static electrical discharges as an intermittent artifact.

Pedants will spot that the first example can be modulated by the position of the magnet room door, rendering the artifact intermittent, while in the second example the propensity for static electrical discharges will persist as long as the source exists, while the humidity remains low, etc. So, yeah, in some ways the distinctions I'm making are subjective. FMRI, like life, is complicated! Still, I'm hopeful that a more practical characterization of artifacts will assist you in differentiating and diagnosing them when it matters: during your experiment. And once you're an expert you will find it easy to comprehend the nuances of temporal behavior, when my artificial distinctions will be all but irrelevant to you.

So, there you have it. I'll be going back to edit the existing posts in this series over the next couple of days. Apologies for any confusion the switch creates.

Thursday, February 16, 2012

Physics for understanding fMRI artifacts: CONTENTS

Figured it might be useful to have some summary/contents pages. I'll do similar admin posts as the other series mature, too. And I will label these pages with "Contents" to make them easier to find via the sidebar.


Part One

An introduction to the series, followed by an introductory video courtesy of Sir Paul Callaghan: What is NMR and how does it work?


Part Two

Further videos explaining the principles of nuclear magnetic resonance - how the intrinsic spin of certain atomic nuclei interacts with applied magnetic fields to yield useful information.


Part Three

Videos showing the anatomy of a miniature scanner, a basic NMR experiment, why shimming is important for NMR (and MRI), how and why a spin echo works, and the relaxation of spins back towards their ground state.


Part Four

Mathematics of oscillations: an introduction to imaginary and complex numbers, and frequency and phase.

Wednesday, February 15, 2012

Common persistent EPI artifacts: Abnormally high N/2 ghosts (2/2)


In the previous post I covered sources of persistent ghosts that arise as a result of some property of the subject, such as the orientation of the subject's head in the magnet. These are what I'm categorizing as subject-dependent effects. In this post I will review the most common sources of persistent ghosts attributable to the scanner, either from an intrinsic property that you might encounter inadvertently, or from mis-setting a parameter in your protocol. As I mentioned last time, I am restricting the discussion to factors that you have some control over as the scanner operator. Ghosts that arise because of a scanner installation error, such as poor gradient eddy current compensation or inaccurate gradient calibration, are issues for your facility physicist and/or your service engineer.


Scanner-dependent conditions:


Rotated read/phase encode axes 

GLOBAL - affects all slices to some extent.

This is an insidious problem that we could categorize as pilot error, except that it's very easily encountered without realizing it. When you set up your slice prescription you are primarily concerned with capturing all those brain regions you need for your experiment. Or you might be concerned with setting a particular slice angle relative to the brain anatomy, e.g. parallel to AC-PC. Now, if the subject's head is precisely aligned such that the read and phase encode axes of your imaging plane are matched perfectly with the gradient set axes (i.e. with the magnet's frame of reference), then for axial slices the readout dimension will be attained using pure X gradient (subject's left-right) while the phase encode dimension uses pure Y gradient (subject's anterior-posterior). (See Note 4 in the post on "Good" coronal and sagittal data for an explanation of why the gradients are established this way, for subject safety/comfort reasons.) But, if the head is twisted slightly, or you're a little sloppy with your slice positioning, then it is quite easy to have a readout gradient that is mostly X with a little bit of Y, and a phase encoding gradient that is mostly Y with a little bit of X. This in-plane rotation ought not be a problem if the X and Y gradients performed equivalently, but they're only similar and not identical. There tend to be small differences in the response time of the gradients, which means that when the scanner tries to drive the read gradient to its desired k-space trajectory, one component (say the X component) can respond faster than the other. This produces a slight mismatch between the target (ideal) k-space trajectory and the trajectory that's actually achieved by the gradients, thereby leading to a source of zigzags that will produce N/2 ghosting.

Now the good news. You've got to rotate the image plane by quite a lot before the ghosting starts to become apparent. It's common to have rotations of 1-2 degrees and these will generate almost no additional ghosting. Once the rotation gets much larger than 5 degrees (depending on the specifics of your scanner) then you might start to see additional ghosting. Below on the left is an ideal prescription, while on the right I've intentionally rotated the image plane by 8 degrees, leading to a small but noticeable increase in ghost level:

(Click to enlarge.)